产品说明
一般描述
Maintains a slight positive pressure in any Plas-Labs glove box. This self regulating purge system controls nitrogen (inert) gas consumption by regulating flow rate. An integral humidity sensor continually monitors interal humidity levels. If the humidity inside the chamber falls below the adjustable setpoint, the controller switches to high gas flow. Once the desired humidity has been restored the controller resumes low gas flow. Basic features include: programmable logic controller, humidity sensor, black material housing, two manually adjustable high and low gas flow knobs, audible alarm, timer, solenoid valve, stainless steel compression fittings for gas line hook-up.
Humidity range: 0-100%
Accuracy: 0.5% for full scale
Setpoint accuracy: 0.1% for full scale
Gas connection: 3.1 mm NPT (
Gas inlet pressure: 4 psi
法律信息
null
基本信息
NACRES | NB.35 |
产品性质
manufacturer/tradename | Plas-Labs 800-CPS/EXP (Euro) |
availability | not available in USA and Canada |
measuring range | resolution: 0.01 |
参数 | 0-20 flow rate (SCFH) |
高度 × 宽度 × 直径 | 7 1/2 in. × 8 1/4 in. × 9.0 in., overall |
Sigma-Aldrich